Operation
Page 44 Manual Bühler 1027
3.4.1
**********************
MODE
EVENT
*TIME
FLOW ANAL. 0-20 mA
**********************
The selected sampling mode is marked with a star.
If the sampling mode shall be changed, press the ENT key and move
the cursor on the sampling mode TIME to create a time-proportional
program.
Confirm the input by pressing the ENT key.
4.12.1
**********************
INTERVAL
00:10
**********************
Enter the sampling interval (hh:mm).
A sample is e.g. extracted every 10 minutes.
5.2.2
**********************
BOTTLE FILLING
*TIME-RELATED
SAMPLE-RELATED
**********************
Select whether the bottle filling shall be
1. time-dependent or
2. depend on a pre-programmed number of
samples
6.2.1
**********************
FILLING TIME PER
BOTTLE
2:00H
**********************
TIME-RELATED bottle filling:
Enter for how long each bottle shall be filled until the distributor
changes to the next bottle.
Example:
Interval: 10 min.
Filling time/bottle: 120 min.
Bottle volume: 2,5 L
That means 12 samples are extracted in 120 min. With a bottle
volume of 2,5 L, the maximum dosing volume per sample is
calculated as follows:
2500:12 200 ml.
5.2.5
**********************
BOTTLE VERSION
*SINGLE BOTTLE
2 x 5;0 L
2 x 10,0 L
**********************
If your sampler is equipped with a distributor, select here the
distributor/bottle variant.
5.2.12
**********************
REPEAT PROGRAM
YES
*NO
**********************
If the program shall only run once, select NO.
Selecting „YES“ means continuous operation (the program is
always restarted).
Remark: In that case, the program will run endlessly. So, please
make sure that the bottles are emptied regularly
according to the programming as otherwise they will be
overfilled.